Please use this identifier to cite or link to this item: http://dspace.sctimst.ac.in/jspui/handle/123456789/9723
Title: EVALUATION OF MATERIALS FOR ARTIFICIAL-HEART VALVES
Authors: BHUVANESHWAR, GS
MURALEEDHARAN, CV
RAMANI, AV
VALIATHAN, MS
Keywords: Materials Science
Issue Date: 1991
Publisher: BULLETIN OF MATERIALS SCIENCE
Citation: 14 ,6;1363-1374
Abstract: The heart of a normal human being beats about 38 million cycles every year. An artificial heart valve, to perform at this rate in the adverse conditions inside the heart for 20 years or more, should be highly wear-resistant with excellent fatigue strength. Thus, the study of mechanical and physical properties of the materials intended for use in artificial valves becomes an inseparable part of the valve development process itself. The physical and mechanical requirements of the materials used in the Chitra heart valve have been evaluated by studying their water absorption, adhesive wear and abrasive properties. The mechanical durability of the device has been assessed by accelerated life cycle testing. The test systems developed for the above are described here. The results show UHMW-PE to be a highly wear-resistant material suitable for the occluder. The accelerated wear tests show that the valve with Haynes-25 alloy cage and UHMW-PE disc has durability in excess of 50 years.
URI: 10.1007/BF02823243
http://dspace.sctimst.ac.in/jspui/handle/123456789/9723
Appears in Collections:Journal Articles

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.