Khambete, NDBrown, BHSmallwood, RH2017-03-102017-03-10200021 ,1;79-8810.1088/0967-3334/21/1/310https://dspace.sctimst.ac.in/handle/123456789/10087In this paper, we have proposed a technique for reducing movement artefacts in impedance pneumography by placing six electrodes at appropriate locations and suitably combining the measurements obtained. The strategy for electrode placement was based on the observation that the electrodes appeared to slide over the rib cage along with the skin, during movement. A volume conductor model of the thoracic cavity was developed and movement artefacts were simulated by shifting the electrodes to a different location on the surface. The impedance changes due to movement in one of the measurements of a 'symmetrical pair' were 180 degrees out of phase with respect to those observed in the other measurement of that pair. However, the impedance changes due to breathing were in phase in both these measurements. Thus, it was possible to reduce movement artefacts by taking a mean of these measurements without affecting the breathing related changes. The six electrodes could be configured into two such symmetrical pairs. The same observation was made in experimental data recorded from human subjects. This indicated that movement artefacts were caused by sliding of electrodes along with the skin and could be reduced by using the six-electrode configuration.Biophysics; Engineering; PhysiologyMovement artefact rejection in impedance pneumography using six strategically placed electrodes