Influence of polyethylene glycol graftings on the in vitro degradation and calcification of bovine pericardium

No Thumbnail Available
Date
1997
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
JOURNAL OF BIOMATERIALS APPLICATIONS
Abstract
Calcification is a frequent cause of the clinical failure of bioprosthetic heart valves fabricated from glutaraldehyde pretreated bovine pericardium (GATBP). This article reports on various chemical techniques for grafting polyethylene glycol (PEG) on bovine pericardium, their biostability, and calcification. The process of calcification profile was studied by in vitro experiments via the incubation of pericardial samples in a metastable solution of calcium phosphate. The calcification profile of PEG-modified bovine pericardium through glutaraldehyde linkages was significantly reduced compared to other methods of grafting. The mechanical property of these PEG-modified tissues after enzyme (collagenase) digestion and calcification were also investigated. PEG grafting of BP via glutaraldehyde or hexamethylene diisocyanate had shown better mechanical stability compared to other grafting methods used.In conclusion, it seems that the surface modification of bovine pericardium through high molecular weight PEGs via glutaraldehyde linkages may provide new ways of controlling tissue biodegradation and calcification.
Description
Keywords
Biocompatibility
Citation
JOURNAL OF BIOMATERIALS APPLICATIONS. 11; 4; 430-452
Collections