EVALUATION OF SOFT-TISSUE RESPONSE TO A POLY(URETHANE UREA)

dc.contributor.authorMOHANTY, M
dc.contributor.authorHUNT, JA
dc.contributor.authorDOHERTY, PJ
dc.contributor.authorANNIS, D
dc.contributor.authorWILLIAMS, DF
dc.date.accessioned2017-03-10T03:26:37Z
dc.date.available2017-03-10T03:26:37Z
dc.date.issued1992
dc.description.abstractVariations in the performance of vascular prostheses constructed of polyurethanes, and some evidence which suggested that these variations could be due not to the properties of the polymer itself, but to differences in the cellular response to the various microstructures of porous polyurethanes require investigation. Experiments were performed to evaluate quantitatively the extent of the cell behaviour adjacent to a series of polyurethane samples. It was shown that, with Biomer(R), a polyurethane urea, the profile of cell behaviour as a function of distance from the implant surface and of time following implantation, the response of cells in general and macrophages in particular, varied considerably with different internal microstructure. This supports the suggestion that the cellular response to different structures and susceptibility to degradation are related.
dc.identifier.citation13 ,10;651-656en_US
dc.identifier.uri10.1016/0142-9612(92)90123-6
dc.identifier.urihttps://dspace.sctimst.ac.in/handle/123456789/9725
dc.publisherBIOMATERIALS
dc.subjectEngineering; Materials Science
dc.titleEVALUATION OF SOFT-TISSUE RESPONSE TO A POLY(URETHANE UREA)