A comparative evaluation of coefficient of friction and mechanical properties of commercially available foley catheters

dc.contributorRamesh, P
dc.contributorJoseph, R
dc.contributorSunny, MC
dc.date.accessioned2012-12-04T11:43:11Z
dc.date.available2012-12-04T11:43:11Z
dc.date.issued2001
dc.description.abstractThree commercially available urinary catheters were evaluated for their coefficient of friction and mechanical properties, which are two important parameters as far as the intended use is considered. The apparent coefficient of friction was measured from the slope of the force versus normal load graph using an in-house fabricated setup. The mechanical properties like ultimate tensile strength, elongation at break and modulus were evaluated before and after aging in synthetic urine. The swelling behavior of the catheter samples was studied in toluene and found to correlate well with the mechanical properties.
dc.identifier.citationJOURNAL OF BIOMATERIALS APPLICATIONS. 15; 4; 344-350en_US
dc.identifier.urihttp://dx.doi.org/10.1106/NCRR-MDV5-FQFG-L7PG
dc.identifier.urihttp://www.ncbi.nlm.nih.gov/pubmed/11336387
dc.identifier.urihttps://dspace.sctimst.ac.in/handle/123456789/12
dc.publisherJOURNAL OF BIOMATERIALS APPLICATIONS
dc.subjectMedical Devices
dc.titleA comparative evaluation of coefficient of friction and mechanical properties of commercially available foley catheters