A sensitive estimation of residual ethylene glycol in ethylene oxide sterilized medical devices by HPLC with electrospray ionization mass spectrometric detection

No Thumbnail Available
Date
2009
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
Abstract
A novel analytical methodology for the estimation of residual ethylene glycol (EG) in ethylene oxide sterilized polymer is reported. The method involves the monitoring of ammonium adduct of EG ions in the presence of 10 mM ammonium acetate buffer and methanol using electrospray ionization liquid chromatography and mass spectrometry (LC-ESI-MS). The method enables the detection and quantification of EG without prior derivatization up to a level of 0.06 mu g/ml. The potentiality of the method is demonstrated by estimating EG in ethylene oxide (EtO) sterilized polyethylene terephthalate fabric used in heart valve sewing ring. The method is simple, rapid and can routinely be used for the quantification of residual EG in EtO sterilized medical devices. (c) 2008 Elsevier B.V. All rights reserved.
Description
Keywords
Sterilization
Citation
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES. 877; 3; 328-332