Preclinical evaluation of hydrogel sealed fluropassivated indigenous vascular prosthesis
No Thumbnail Available
Date
2018-03
Journal Title
Journal ISSN
Volume Title
Publisher
Indian Journal of Medical Research
Abstract
Background & objectives: Polyethylene terephthalate (PET) graft, designed and developed at our
institute for vascular reconstruction, is porous to promote optimal incorporation and neointima
formation, requiring pre-clotting or biomodification by sealing the pores before implantation. The
objective of this study was to characterize, test and perform preclinical evaluation of hydrogel
(alginate dialdehyde cross-linked gelatin) sealed fluoropassivated PET vascular prosthesis in pig model,
so as to avoid pre-clotting, for its safety and efficacy before employing the indigenous and less expensive
graft for clinical use.
Methods: Hydrogel sealed, fluoropassivated PET vascular prosthesis were tested for haemocompatibility
and toxicity followed by small animal toxicology tests and in vivo experiments in pigs receiving
implantation at thoracic aorta. All 33 animals received test as well as control grafts with a plan for
phased explantation at 2, 12 and 26 weeks. All animals underwent completion angiogram at the end of
procedure as well as before graft explantation.
Results: Haemocompatibility tests for haemolysis and toxicity tests showed no adverse events in tested
mice and rabbits. Completion angiogram showed intact anastamosis and patent graft in each animal
in post-operative period and at explantation. Gross and histopathological examination showed wellencapsulated
grafts, clean glistening neointima and no evidence of thrombus in both test and control
grafts.
Interpretation & conclusions: Hydrogel sealed, fluoropassivated PET vascular prosthesis was found
non-toxic, haemocompatible and remained patent in in vivo studies at planned intervals.
Description
Keywords
Aortic reconstruction - flouropassivated polyester graft - hydrogel sealing - in vivo large animal trial - polyethylene terephthalate vascular prosthesis
Citation
Unnikrishnan M, Umashankar PR , Viswanathan S, Savlania A, Joseph R, Muraleedharan CV, Agrawal V, Shenoy SJ, Krishnan LK Mohanan PV, Sabareeswaran A. Preclinical evaluation of hydrogel sealed fluropassivated indigenous vascular prosthesis. Indian Journal of Medical Research. 2017;146(5):646-53