Polyethylene glycol (PEG) modified bovine pericardium as a biomaterial: A comparative study on immunogenicity

No Thumbnail Available
Date
1998
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
JOURNAL OF BIOMATERIALS APPLICATIONS
Abstract
Bioprosthetic heart valves made from glutaraldehyde (GA)-fixed porcine aortic valves or bovine pericardium (BP) are having some advantages over mechanical valves. However, their durability is low due to the calcification and immunological rejection. Study on immunogenicity is an important part in understanding the biocompatibility of materials. Polyethylene glycol (PEG) on pericardium can control biodegradation and calcification. Also, PEG exhibits low immunogenicity. We have studied the complement activation potential and the contribution of complement factors (biologic factors) on the calcification of PEG grafted pericardium samples and compared with standard (control) glutaraldehyde-treated pericardium samples. PEG-grafted BP activated using GA and carbodiimide (EDC) could be selected for further studies since complement activation and calcification observed on these samples has been relatively low.
Description
Keywords
Biocompatibility
Citation
JOURNAL OF BIOMATERIALS APPLICATIONS. 13; 2; 158-165
Collections